Signal Conditioning Unit
The RM-044 has been designed as a low cost, stand-alone unit to provide signal conditioning for strain gauge type transducers such as Pressure Transducers and Load Cells. It can also be factory configured for ½ bridge and ¼ bridge 120 ohm and 350 ohm strain gauges. Gain selection and gross offset correction are all internally link selectable. The electronics are housed in a 100 x 115 x 20 mm plastic housing. Four, three pin connectors are provided for signal, transducer and power connections. The RM-044 is normally supplied to provide + 10 volts output for the full range of the transducer, but the output will drive over the full range of ± 10 volts. The output is also configured to provide 4 to 20mA signals if required.
SPECIFICATIONS
Power Supply: +24 Volt Nominal. (18 To 30 V).
Transducer Excitation: +1.3 -> +10 VOLT @ 50 mA MAX. (or +/- 15 Volt unregulated @ 15 mA Max. Link selectable).
Excitation Stability: 0.02 % / °C
Input Impedance: > 100 M ohm (Differential)
CMRR: 100 dB @ 50 Hz Linearity: 0.01 %
Gain Range: 120 TO 3200
Frequency Response: 20 Hz Or 1 kHz 4 Pole Max Flat (Factory Configured)
Outputs: ± 10 VOLTS @ 5 mA MAX. 4 to 20 mA ( 500 ohm MAX ).
Output noise: 1 mV RMS . Environmental 0 TO 70 °C 90% R.H. N/C.
Physical: 100 x 115 x 20 mm (LWH) |
||||||||||||||
|
||||||||||||||
|
||||||||||||||